optional settings
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 104 of 213
[8.A2]
OP44
Locking and unlocking the test mode
OFF = test mode is locked / not accessible
1 ... 94 = test mode is unlocked / accessible up to the selected test
profile number Standard: 24
selected test number = 5: The test profiles 1-5 can be selected, the
test profiles 6 – 94 can not be selected
OP45
Report mode display „index“
"
0 = Display automatically switches through (Default)
1 = Display permanently in MET
2 = Display permanently in KJ
3 = Display permanently in Watt
After total switch off the default value 0 will be valid again.
OP46
Report mode display „elevation“
$
in „Profile“ and „Test“ mode
0 = Display automatically switches through (Default)
1 = Display permanently in % or degree (°), depending on OP14
2 = Display permanently in „step“
After total switch off the default value 0 will be valid again.
OP47
Sustain values in display resp. automatic
„Reset“
OFF = Display values are deleted after pressing START again or
automatically after 2 minutes after having pressed STOP (Default)
ON = Display values will be continued (added) after pressing START
again and will not be automatically deleted by pressing STOP.
Display values can only be deleted by pressing the STOP key twice
(time, distance, energy).
OP48
Count-Down of program-step
OFF = The time-display counts up each program-step
ON = The time-display counts down each program-step
OP52
Output interval for printer protocol
By entering a value between 0 and 100 the output interval is set in
seconds for a printer directly connected to the treadmill. Standard: 60
(= Printout of all values once a minute). The value 0 disables the
printing out of single values, but not the printing of headers and end
results (UKK).
OP53
Language settings for printer
protocol
Select the language for printouts on a printer directly connected
to the treadmill. You can choose between six languages. Both
the protocol printout as well as the test result and training
recommendation of the UKK 2 km walk test are printed in the
selected language
.
EnGL = Englisch (standard)
SPAn = Spanish
GErM = German
POrt = Portuguese
FrEn = French
HUnG = Hungarian
For a correct printout the connected printer must be compatible with
PCL printer language. For special characters the ISO 8859-1 (Latin-1)
font is used.