10
80834 REV ED
EN
EN
2.9 Symbol Identifications on labeling and unit
1
2
3
4
2X T5A 250V
5
6
7
SN
8
9
10
11
#
12
Rx Only
13
Distributed
by
14
15
I
O
~
1. Type BF applied part
2. Power ON
3. Power OFF
4. Alternating current
5. Fuse
6. Footswitch
7. Transducer output
8. Serial Number
9. Instructions for Use
10. Caution, Consult Accompanying Documents.
11. Tested by TUV Product Service to UL60601-1: 2003,
CSA 22.2 No. 601.1-M90
12. Quantity symbol: Numeral in symbol (in place of #)
indicates the quantity of units in package.
13. Caution: Federal (US) Law restricts this device to sale
by or on the order of a physician.
14 Distributed by
15. Reference Directive 2002/96/EC.
(O-Ring)
4
2
3
1
Section 3.0 Assembling the Unit
1. Disinfect and sterilize transducer, probe, cleaning stylet, and wrench according to the instructions
in Sections 5.2, 5.3, and 5.4. Note: The stylet must be removed from within the probe prior to
disinfection and sterilization.
2. Disconnect the transducer cover (nose cone) from transducer body to expose the probe
attachment screw threads
3. Firmly wipe the threads in the transducer and the mating screw end of the probe with a sterile
4” x 4” gauze pad to remove any moisture and to permit a firm seating.
4. Check to ensure that the Probe set parts are in the correct order before attaching probe to
transducer.
1. Floating (outer) Probe
2. Return
Spring
3. Fixed (Inner) Probe
4. Free Mass Ring
First, finger-tighten the Fixed (inner) Probe to the Transducer. Then use the wrench to be sure that
the Fixed Probe is screwed tightly onto the Transducer.
CAUTION: Probe must be secured tightly in transducer to assure good coupling. Loosening of
the probe during operation will decrease or halt lithotripsy action of the probe and may cause
damage.
Caution: If the probe is not properly assembled, seated or tightened into the transducer, it
will not perform properly. Check to ensure that the probe is assembled and seated correctly
prior to use.
If the probe does not appear to work effectively, use a cleaning stylet to make sure it
is not obstructed by debris.
Do not use a damaged probe. Probes (
REF
CW-USLRBP) are supplied non-sterile* and must
be sterilized prior to use. CyberWand probes are designed as single use devices. DO NOT
REUSE. BREAKAGE MAY OCCUR.
* Sterile 3-pack Probe Sets (
REF
CW-RBPBX) are available, sold separately.