Section 6.0 Operating and storage environment
Operating environment
Ambient temperature
10 – 40°C
Relative humidity
30% – 70%
Air Pressure
70 – 106 kPa
Storage environment
Ambient temperature
-25 – +55°C
Relative humidity
10% – 85%
Air Pressure
70 – 106 kPa
Section 7.0 Preventive Maintenance
7.1 General
Checks
The operator should ensure the generator and all cables and all components on transducer are
undamaged prior to beginning any procedure.
7.2 Periodic Safety Checks
The safety checks should be performed at least every 24 months.
The inspection can only be
performed by a person who has adequate training. knowledge, and practical experience. All the
equipment used in safety check must be calibrated. Gyrus ACMI recommends returning the unit for
safety inspection or contact your local representative for information about maintenance contracts.
Section 8.0 Troubleshooting
8.1 Troubleshooting
Irregular Description
Possible cause
Solution
Power indicator LED is not
illuminated.
The power cord is not securely
plugged into a receptacle
Connect power cord into receptacle.
Note: Ensure the power switch is at OFF
position before plugging in.
The power cord is not securely
plugged into generator
Connect power cord into generator.
Note: Ensure the power switch is at OFF
position before plugging in.
Power switch is not turned ON
Turn the switch ON.
The fuse has blown
Replace fuse as described in 8.2 Fuse
Replacement below.
Fault Indicator light is
illuminated.
Footswitch is not plugged into
generator
Plug footswitch into generator.
Note: Ensure the power switch is at OFF
position before plugging in.
Transducer is not plugged into
generator
Plug transducer into generator.
Note: Ensure the power switch is at OFF
position before plugging in.
Poor performance or low
power
• Mis-tuned probe
• Loose Probe
• Re-tune system (Section 3.1).
• Retighten probe and re-tune system.
8.2 Fuse
Replacement
Warning:
Ensure that the power cord is disconnected from the instrument before opening the fuse holder.
1. Remove the fuse holder with a flat screwdriver, to expose fuse
2. Replace the blown fuse with a new fuse (TYPE T 250V 5A).
3. Replace the fuse holder securely into the generator.