Install the Funnel Rack
the 16-outlet funnel rack accommodates large and variable collection volume in external
high-capacity containers. the funnel rack is installed directly on the collector tray in front of
the system and replaces the standard racks.
to install the funnel rack:
1. use a 2.5 mm allen wrench to remove the two cylindrical-head screws from the front
positioning guide.
2. remove the front positioning guide from the tray.
3. Place the funnel rack on the tray and fix it with the two cylindrical-head screws previously
removed. tighten the two screws with the 2.5 mm allen wrench.
4. Connect the 16 outlet tubing and securely tighten the black nuts.
Figure 46
funnel rack
Figure 47
funnel rack installation
front Guide Screws
front Guide Screws