Install the Racks
PLC 2050/2250/2500 Purification Systems accommodate three racks. the standard, supplied
rack set (Pn 21040019) consists of three SS racks with 192 tubes, 180 mm high with a total
collection volume of 6.1 L.
NOTE
Remove the black cable tie which
holds the collector arm during
transport by manually loosening it.
Keep this cable tie, it is reusable.
to install the racks:
1. Place the tubes in the rack holes, and then
place the racks on the collector plate.
2. Locate the middle slot in the rear of
the rack. Slide the slot over the back
positioning guide mounted on the collector
plate.
3. fit the middle slot on the front of the rack over the positioning guide in front of the
collector plate.
NOTE
When using short racks, use the additional positioning guide supplied and place it in appropriate
holes.
Figure 33
Collector arm Cable tie
Dispense
arm
Cable
tie
Figure 34
rack installation
Figure 35
Short racks Guide
Back Guide
rack
rack Slot
front
Guide
Back Guide for
Short racks