to install the clamp:
1. Slide the rod horizontally into the black holder.
tighten the two cylindrical-head screws with the
3 mm allen wrench supplied.
2. to install the column, loosen the wing nut, and
then slide the column inside the clamp. Securely
tighten the clamp.
3. Connect the column inlet to the loop injection
valve outlet (manual or automatic).
●
Port 5 for the automatic loop injection valve
●
Port 3 for the manual loop injection valve
4. Connect the column outlet to the bulkhead
union detector inlet.
NOTE
If additional valves are installed, such as the
automatic backflush valve (PN 21040005 for PLC
2050; PN 21040006 for PLC 2250 and PLC 2500)
or manual column switching valve (PN 21040013
for PLC 2050; PN 21040014 for PLC 2250 and
2500), proceed as indicated on the system.
5. use a 3/8” spanner to tighten the nuts and crimp
the ferrules.
Column Connections to the Automatic Loop Injection Valve
COMPONENT
DESCRIPTION
COMPONENT
DESCRIPTION
a
Bulkhead union detector inlet
e
automatic loop injection valve
B
Column
f
Column tubing, from column outlet
to bulkhead union detector inlet
C
Column outlet/inlet
G
Column tubing, from automatic
loop injection valve (Port 5) to
column inlet
D
Column outlet/inlet
H
Sample loop
A
B
C
D
E
F
G
G
H
Figure 32
Column Connections to automatic
Loop injection Valve
Содержание PLC 2050
Страница 1: ...PLC 2050 2250 2500 Purification Systems User s Guide...
Страница 6: ...PLC 2050 2250 2500 Purification Systems USER S GUIDE...
Страница 34: ......
Страница 50: ......
Страница 58: ......
Страница 70: ......
Страница 78: ......
Страница 88: ......
Страница 102: ......
Страница 112: ......
Страница 130: ......
Страница 131: ......
Страница 132: ...www gilson com ContactUs LT380160 01...