User Manual
Genevac HT-4X Evaporation System
04-7415 Issue 1-16 August
2014
53
Technical Data
Specifications
Evaporator
Max rotor speed
1745 rpm
Max load per swing
1.5 kg
Max operational imbalance
85 g
Dimensions
( w x d x h)
590 x 640 x 980 mm
Approx weight
( varies with build options)
118 kg
Height include allowance for lid opening
Vacuum pump (remote)
Type
Oil free scroll
Ultimate system vacuum
< 0.4 mbar
Dimensions
(w x d x h)
530 x 305 x 398 mm
Weight
28 kg
Vacuum hose / control cable length
3 m
Condenser (integral)
Type
Single stage vapour compression
Refrigerant
R404a
Ultimate low temperature
-50°C
Storage / transportation environment
Ambient temperature
0°C to 40°C
(-10°C permissible during transportation)
Relative humidity 10% to 80% non-condensing
Store upright
Operational environment
Ambient temperature
15°C to 30°C
Relative humidity 10% to 80% non-condensing
Altitude
Sea level to 1600 m
Ventilation air-gap
50 mm
Ingress protection rating
IP30
The evaporator is designed for use in a
pollution degree 2 environment (normally only
no- conductive pollution occurs).
Emissions
Typical noise level is 69 dB (A) at one metre
from the evaporator during normal operation.
For the purpose of air conditioning requirement
calculations, all power consumed by the
system is dissipated as heat.
Clear PVC hose (ID 13 mm) is supplied to
connect the evaporator to a laboratory fume
extraction system.
Inert gas supply (IGP option only)
Max pressure
8 bar g (9 bar absolute)
Min pressure
3 bar g (4 bar absolute)
Hose length
2.5 m
Connector type
3/8” BSP female
Electrical
Supply
208 V 60 Hz
or 230 V 50 Hz
Power consumption
2.3 kW
Add 200 W for high power lamp option
The power cable supplied is 2 m in length and
appropriate to the region to which the unit is
delivered.
Note:
To prevent nuisance tripping, the mains
power supply to the evaporator should be fitted
with suitably rated type “D” (or equivalent)
mains circuit breaker.