
Genevac HT-4X Evaporation System
User Manual
18
04-7415 Issue 1-16 August
2014
Run parameters
The following pages describe the function of each run parameter in turn and show the range
of setting options available. Instructions for setting the parameter are also given.
Run Data Protection
The run profile can be protected against accidental alteration. Set the
Run Data Protection
field to
No
before attempting to edit an existing run profile or set to
Yes
to prevent accidental
changes being made to the run profile. For additional security, a password may also be set
to protect against unauthorised changes.
Next Run Number
Runs can be linked together, enter the number of the run that requires to be linked and it will
commence when the current run finishes. Any number of runs can be linked together in this
way.
If no link is required, enter 0 (default setting) and the system will stop when the current run
ends.
Note:
A never-ending chain will be created if the last run of a chain in linked back to the
first, or if a run is linked to itself. If such a chain is created the only way to stop the
evaporator is to change the setting or to manually press the
STOP
button.
Run Name
Enter a name (up to 20 characters) which the user can recognise to identify the run profile.
Text entered on the keyboard appears in the lower section of the screen. Press the return
key to enter the text into the
Run Name
field.