Genevac HT-4X Evaporation System
User Manual
48
04-7415 Issue 1-16 August
2014
Troubleshooting
Audible alerts
Notification
Button acknowledge
Short beep
Door lock acknowledge
Short beep
Illegal button press
Short double beep
Evaporator stopped
Three short beeps when the system comes to rest. Repeats
every minute until the door is opened
Critical error
Three short beeps when the system comes to rest. Repeats
every minute until the door is opened
Visual alerts
Notification
System not ready
Message appears at bottom of the
Run log
screen
Non critical error
Descriptive message in
ALARMS
field of
Run log
screen
Critical error
Descriptive message appears and flashes red in
ALARMS
field
of
Run log
screen
Faults and errors
In the unlikely event that an error occurs that that could affect the physical condition of the
evaporator or its ability to safeguard the integrity of the samples (a critical error) the system
will stop automatically. A description of the problem appears in the
ALARMS
field of the
Run log
screen.
If a non-critical error occurs, the system continues to operate (possibly with reduced
efficiency) and a description of the problem is shown in the
ALARMS
field. An example of a
non-critical error could be the failure of an IR lamp, the evaporator would continue to operate
without compromising the safety of samples, but the evaporation rate would slow down.
Clearing Errors
The evaporator will not allow a run to start while an error is displayed. To rectify the cause,
follow the advice on
Rectifying faults and errors
given in the tables on the following pages.
Once the problem is resolved clear the error by switching the evaporator off, wait for ten
seconds then switch on again.