Genevac HT-4X Evaporation System
User Manual
24
04-7415 Issue 1-16 August
2014
Heat-Off Option
Select from the options:
Elapsed Time
Manually enter a time for the
Heat-Off Option
, the evaporator turns off the IR lamps
when the time elapses and continues to run without heat input until the end-of-run time
elapses. By entering a longer time in the
End-of Run Option
, thermally sensitive
samples can be protected as they approach dryness while ensuring that all samples are
dried completely.
Heat Flow
The system monitors evaporative cooling by looking at the demand for heat input
required to maintain the sample holders at control temperature. When sample dryness is
detected, the system automatically turns off the IR lamps and performs a final drying
stage before stopping the run.
Sample Detection
The SampleGuard™ system monitors the temperature of a selected sample. It detects
dryness when evaporative cooling ceases, turns off the IR lamps and then stops the run.
The use of two temperature probes is necessary for
Sample Detection
. See
SampleGuard
for details.
End-of-Run option
Only applicable if
Heat-Off Option
is set to
Elapsed Time
; manually enter a time for
End-of-Run Option
.
End of run defrost and drain
Applicable to
Auto Defrost and Drain
systems only, select yes to automatically defrost and
drain the condenser at the end of the run (or end of the stage if one or more runs are linked).