Genevac HT-4X Evaporation System
User Manual
38
04-7415 Issue 1-16 August
2014
Crystallisation
For evaporators equipped with the optional eXalt crystallisation package, the
Crystallise
option is accessed from
Pressure Control Regime
in the
Run Data
screen.
Selecting
Crystallise
allows the user to enter control parameters for pressure reduction
cycles. Select
Crystallise
and enter values for:
Pressure reduction (mbar)
Pressure reduction time (MM)
Atmosphere Vent Time (MM)
The software feature is designed for use with an array of baffles which are supplied in the
eXalt toolkit. When fitted to vials, the baffles control the evaporation rate for each individual
vial within a sample holder. This allows a wide range of solvents to be evaporated at the
same time and at the same slow rate.