Genevac HT-4X Evaporation System
User Manual
42
04-7415 Issue 1-16 August
2014
Example 2. Fast Lyophilisation of HPLC Fractions
Method consists of three runs linked together to form a multi stage run, with settings as
follows:
Stage 1
Ramp & remove
organic phase
Stage 2
Cool
the water
Stage 3
Freeze
the water
Stage 4
Lyophilise
the water
Run No.
1
2
3
4
Run Data Protection
N
N
N
N
Next Run Number (0 for none)
2
3
4
0
Run Name
Fractions Lyo 1
Fractions Lyo 2
Fractions Lyo 3 Fractions Lyo 4
Sample Holder Type
Your holder
Your holder
Your holder
Your holder
Sample Volume (ml)
Your volume
Your volume
Your volume
Your volume
Solvent
MeCN or MeOH &
H20
MeCN or MeOH &
H20
MeCN or MeOH &
H20
MeCN or MeOH &
H20
SampleGuard Control Temp (°C)
40
0
0
40
SampleGuard Control Channel
1
1
1
1
CoolHeat Enable Pressure (mbar)
50
50
50
50
Chamber Temperature control (°c)
preheat to 25
preheat to 0
preheat to 0
preheat to 30
Rotor Speed
High
High
Low
Low
Heat-Off Option
Elapsed time
Elapsed time
Elapsed time
Elapsed time
Heat-Off Elapsed Time (HH:MM)
ask for time
0:00
00:00
05:00
End-of-Run Option
Elapsed time
Elapsed time
Elapsed time
Elapsed time
End-of-Run Elapsed Time (MM:MM)
ask for time
0:30
01:00
10:00
Auto Defrost & Drain
Y
*
N
N
N
Pressure Control Regime
Variable Dri-Pure Controlled pressure Full vacuum
Full vacuum
Dri-Pure Start Pressure (mbar)
175
-
-
-
Dri-Pure End Pressure (mbar)
40 for MeCN
45 for MeOH
-
-
-
Dri-Pure Ramp Time (MM)_
20
-
-
-
Final Control Pressure (mbar)
40 for MeCN
45 for MeOH
8
-
-
* Available on
LyoSpeed
systems only. For non-LyoSpeed systems, defrost and drain
manually before moving to next stage.