User Manual
Genevac HT-4X Evaporation System
04-7415 Issue 1-16 August
2014
9
To start the system, press the green
START
button on the evaporator keypad.
The vacuum pump starts and the
SELECT RUN
menu appears on the
evaporator LCD screen.
The condenser also powers up and begins to cool down to its operating temperature. The
system will not allow an evaporation run to commence until the condenser reaches 0°C; if an
attempt is made to start, the run will commence once operational temperature has been
reached.
Note:
For HCl systems, an additional pump warm-up sequence must complete before the
system will allow an evaporation run to commence (this prevents the possibility of solvent
condensation inside the pump). The warm-up time remaining is shown on the screen.
To power down, enter the
Select Run
screen and press the
STOP
key. Press Y on the
evaporator keypad when prompted:
Note:
If a pump purge sequence begins, allow this to finish before switching off the
evaporator power switch.