SampleGuard Control Temperature
Enter a temperature value up to +70°C. The system heats the sample holders to this
temperature.
Caution:
Risk to Sample Integrity; consider the possible effects on the dried
sample when selecting a safe control temperature.
SampleGuard Control Channel
Channel 1 is the default setting. The option to change to channel 2 is provided only to allow
continued operation should a fault arise with channel 1 or its associated SampleGuard
probe.
CoolHeat Enable Pressure
Enter a pressure value between 5 and 400 mbar. The infra red lamps do not come on until
the
CoolHeat Enable Pressure
is reached. This prevents a build up of heat that would
cause sudden uncontrolled boiling (or bumping) of volatile solvents as the chamber pressure
reduces
.