ORTHORALIX 9200
Service Manual
B-19
(06.0)E
Gendex Dental Systems 4519-190-71193-7
B.5.12
Mounting the column covers and the
quick reference guide
To mount the covers of the column proceed as follows:
1.
Switch on the system and raise the Z-carriage to the highest
position.
Do not move the Z-carriage along the column
when the system is off:doing so it could damage
the PC6 board
2.
Mount the bottom front cover (a) of the column by means
of four screws (b).
Be sure that the cable retainer of the X-ray command cable
is in the right position.
3.
Mount the four covers (g) on the four column wall brackets,
by means of two screws (h) for each one.
4.
Mount the right lateral cover (e) of the column by means of
seven screws (f).
5.
Mount the left lateral cover (c) of the column by means of
seven screws (d).
Be sure that the mains cable and the remote control cable
pass through the correct cable inlets.
6.
With the system are available two Quick Reference Guide
labels, one for the Orthoralix 9200/Plus and the other one
for the Orthoralix 9200/Plus/DDE; select the correct one and
install it on the right cover of the column, so that the lower
edge of the label is at about 1 m from the floor.
(03.0)E
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER