ORTHORALIX 9200
Service Manual
F-44
Gendex Dental Systems 4519-190-71193-7
F.5.6
Replacing the tubehead
For the replacement of the tubehead proceed as follows:
1.
Remove the front cover (a) of the tubehead by unscrewing
two screws (b) and by disengaging the spring latches from
the rear cover (c).
2.
Remove the rear cover (c) by unscrewing the four Allen
screws (d).
3.
Disconnect the wires from the terminal block 600 of the
tubehead.
4.
Disconnect the ground cable from the tubehead.
5.
Disconnect connector X50 from the collimator.
6.
Remove the collimator assembly from the tubehead, by
unscrewing the four screws (g).
7.
Remove the tubehead (e) by unscrewing four hexagonal
head screws (f).
8.
Proceed backwards for remounting.
(03.0)E