ORTHORALIX 9200
Service Manual
F-41
Gendex Dental Systems 4519-190-71193-7
For the replacement of the R drive belt proceed as follows:
1.
Loosen the dowel (a) and by turning the eccentric (b) with
an open wrench reduce the tension of the belt.
2.
Remove the screws (c) with plates and replace the belt (d).
3.
Ensure that the plates are properly accommodated on the
teeth of the new belt, then fix them with screws (c).
4.
For the adjustment of the belt tension refer to paragraph
Tension adjustment of the drive belts
of this Section F.
(03.0)E