ORTHORALIX 9200
Service Manual
B-15
Gendex Dental Systems 4519-190-71193-7
B.5.9
Installing the remote control box
A remote control box is available for the Orthoralix 9200 system,
to be installed on the wall.
A 20 m. cable is supplied, to connect the control box to the Z-
carriage.
To mount the remote control box proceed as follows:
1.
Feed the connecting cable through the cable inlet at the
rear of the column.
2.
Check that the cable retainer (c) is at the end of the coiled
cable.
3.
Plug the coiled end of the connecting cable into socket X65
of the Z-carriage.
4.
Place the remote control box on the desired installation site
on the wall and mark the fixing holes.
5.
Drill the holes and fit the expansors.
6.
Feed the connecting cable through the hole (e) of the remote
control box.
7.
Connect (f) to (g) (X2) and (h) to (i) (X1).
8.
Fit the cable retainer (l) on the connecting cable and push
it into the hole (e).
9.
Fix the remote box to the wall with two screws.
10.
Connect the exposure handswitch to socket X1 of the remote
control box and hang it on the hook provided.
In order to comply with Federal regulations this remote box
must be installed so the technique factors are visible to the
operator from the remote location.
Cable retainer
Right position
(03.0)E
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER