ORTHORALIX 9200
Service Manual
B-18
Gendex Dental Systems 4519-190-71193-7
Connecting to a networking device (dynamic IP address)
1.
Connect the Ethernet cable
from an available Ethernet port
on the networking device to
the Ethernet port at the base
of the 9200 column.
2.
Install the imaging application software on the PC.
3.
Insert the GxPicture CD-ROM into the PC’s CD-ROM drive
and install the ActiveX driver for the Orthoralix 9200. Refer
to the GxPicture Quick Installation Guide for more informa-
tion regarding this software. An icon will appear on your
PC’s desktop that can be used to launch the Configuration
Manager.
4.
Insert the supplied CD-ROM and install the necessary
sensor.dat file. An icon for the Orthoralix 9200 will appear
in the system tray near the clock on your PC’s desktop. This
icon can be used to launch the Orthoralix 9200 Settings
Manager.
5.
Turn off the Orthoralix 9200. If you only have one machine to
install, please proceed to Step 10. Continue to Step 6 if you
have more than one machine to install or already installed.
6.
Press and hold the No Radiation
and Patient Selection
buttons on the 9200 control panel. While holding them, turn
on the power. This will start Service Mode.
7.
Press the Anterior Carriage Movement
button several
times until the IP address appears on the display. The default
IP address of 10.10.10.1 is set in the factory. If you have more
than one 9200 or need to change the IP address for some
other reason, use the following buttons:
kV Decrease and Increase:
Use these to change the first
set of numbers (ex. XXX.16.137.180
mA Decrease and Increase:
Use these to change the
second set of numbers (ex. 172.XXX.137.180)
s Increase or Decrease:
Use these to change the third set
of numbers (ex. 172.16.XXX.180)
Projection Browsing:
Use these to change the fourth set
of numbers (ex. 172.16.137.XXX)
Note: If you have more than one 9200, you only need to
change the last set of numbers in the IP address to some-
thing between 2 and 9 (i.e. 10.10.10.7).
When you are finished making any changes,press and hold
the No Radiation button to save the new IP address.
8.
Press the Anterior Carriage Movement
button until the
Subnet Mask is displayed. The default Subnet Mask is If you
need to change the Subnet Mask, use the same buttons
that were used when changing the IP address. When you
are finished, press and hold the No Radiation
button to
save the new Subnet Mask.
9.
Turn the 9200 off and then on.
10. Click the DDE Pan icon in the system tray.
11. Click the Configure tab, make sure Auto is selected, and then
click
Insert
.
12. Press the Reset button on the 9200 control panel.
If any of the network information does not match between the
PC and the 9200, perform the following steps:
1.
Click Windows
Start
and then
Run
.
2.
Type
cmd
and click
OK
.
3.
Type
ping x.x.x.x
(where x.x.x.x is the IP address) and press
Enter.
4.
Repeat Steps 6 through 10 of the installation instructions.
Note: Whenever you replace the PC104 board within the
9200, you must restart the PC is is connected to.
Icon Status
The following descriptions indicate the various states of the Set-
tings Manager icon in the system tray.
Blue and Green:
There is a good connection between the
PC and the DDE.
Red:
That is no connection between the PC and the DDE.
Verify your settings and redo the installation instructions.
Yellow X:
There is a connection, but another user is in control
of the DDE. To change this, go to the Orthoralix DDe Setup
Screen, click the Multiroom tab, and then click Take Control.
It may take a few moments before control is transferred.
Red X:
There is a connection but the Sensor.dat file is miss-
ing. Re-install the file from the Installation CD-ROM.
(06.0)E
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER