ORTHORALIX 9200
Service Manual
B-9
Gendex Dental Systems 4519-190-71193-7
In order to do not damage the mains switch (it is located
under the patient handgrip) put the Z-carriage momentarily
on the two spacers located under the patient handgrip: the
Z-carriage must be kept in position by one person.
2.
Use three men to raise the Z-carriage and fit it on to the two
driving pins (a).
3.
Connect free cable connector X59 to the relevant socket
located on the sliding block, to power the Z-carriage from
the mains supply.
4.
Connect the connectors X72A with X72B and X71A with X71B
to power the motor of the column.
5.
For Orthoralix 9200 DDE only: feed the connector X11 into
the Z/Carriage.
Plastic foam
6.
The men at the front keep the Z-carriage pushed against the
column.
One man at the sides fixes the Z-carriage to the screw studs
(c) of the sliding block, by means of the four ring lock-wash-
ers and nuts supplied.
Apply max. torque 15 Nm.
(03.0)E