6.
Remove the connector X37 from the host computer board
PC1 and disconnect the connector connected to the red
and black wires (power supply for the lateral laser). Then,
connect the connector X37 coming from the ceph arm to
the PC1 and connect also the connector to the new red and
black wires.
7.
Fix the ceph arm (c) to the Z-carriage by means of the four
hexagonal head screws (M8x25) and rings supplied.
8.
Remove the panel (h) from the right-hand cover (a).
9.
Replace the right-hand cover (a) of the Z-carriage.
10.
Mount the ear plug support (d) into the cephalostat by means
of the two screws (e). The ear plug support has two vertical
mounting positions: please ask to the user which position
in more convenient.
11.
Mount the nasion rest (f) into the pin and fix it with knob (g).
12.
Install the toroid supplied with the ceph arm on the wires of
the connector X8 of the PA7 board.
13.
Install the toroid supplied with the ceph arm on the wires of
the connectors X10 and X15 (one toroid for both connectors)
of the PA7 board.
Do not install the top cover of the ceph arm till all the ceph
alignments are done (see Section C and Section F of this
Service Manual).
B.5.5
Mounting the front panel with mirror
The front panel with mirror is supplied disassembled from the
Z-carriage.
To mount it proceed as follows:
1.
Turn the rotating arm to the 90° position and move it in order
to have adequate access to the electronic rack.
2.
Fit the top of the front panel with mirror into the slot (b) of
the Z-carriage.
3.
Connect the two centering lights (laser projectors) to con-
nector X31 of the system control board PC1.
4.
Check the proper insertion of cover (d) into the springs (e).
5.
Lastly, close the front panel and fix it by means of the two
Allen screws M5x15 (c) provided.