ORTHORALIX 9200
Service Manual
B-22
(06.0)E
Gendex Dental Systems 4519-190-71193-7
B.5.14
Installing the soft tissues filter
The soft tissues filter must be mounted after the centering of the
ceph X-ray field as described in the section C and F of this manual.
For the mounting of the soft tissues filter see the following pro-
cedure:
1.
Put the position adjusting lever (a) in the vertical position: in
this way the soft tissues filter is corresponding to the posi-
tion 2 (see the reference label placed on the tubehead front
cover).
2.
Mount the soft tissues filter (c) on the position lever using
the two screws (b), so that the distance between the middle
position of the filter and the opposite side of the lead plate
is:
t NNGPSUIF9DNWFSUJDBMBTZNNFUSJD
t NNGPSUIF9DNIPSJ[POUBMBTZNNFUSJDPS
X 24 cm.
Do not change the position of the adjusting lever during the
measurement.
In the Orthoralix 9200 DDE Ceph, the soft tissues filter is not
present: the filtration for the soft tissues is done during the
image acquisition.
For the USA versions, the distance between the middle position
of the filter and the opposite side of the lead plate is:
t NNGPSUIFwYwWFSUJDBMBTZNNFUSJD
t NNGPSUIFwYwIPSJ[POUBMBTZNFUSJD