ORTHORALIX 9200
Service Manual
F-27
Gendex Dental Systems 4519-190-71193-7
F.4.3.10
Converter input read
8JUIUIJTTFSWJDFQSPHSBNZPVDBOWJTVBMJ[FUIFTUBUVTPGUIFIBSE
-
XBSFJOQVUTPGUIFDPOWFSUFSCPBSE12XJUIGPVSIFYBEFDJNBM
numbers.
*OUIJTXBZZPVDBODIFDLXIFUIFSUIF1)CPBSETBSFSFDPHOJ[JOH
the right inputs signals.
The correspondence between the hexadecimal numbers and the
status of the hardware inputs is shown in page F-73.
F.4.3.11
Pan and Ceph PH3 input read
This service program is applicable only for Orthoralix 9200 DDE
versions.
8JUIUIJTTFSWJDFQSPHSBNZPVDBOWJTVBMJ[FUIFTUBUVTPGUIF
hardware inputs of the panoramic PH3 and ceph PH3 boards,
with four hexadecimal numbers.
In this way you can check whether the converter board is recog-
OJ[JOHUIFSJHIUJOQVUTTJHOBMT
The correspondence between the hexadecimal numbers and the
status of the hardware inputs is shown in page F-74.
The following message will appear on the display:
PH3 PAN : CEPH READ
X XX: X XX
The first two numbers are related to the panoramic PH3 board
and the second two are related to the ceph PH3 board.
F.4.3.12
Exposure counter
#ZTFMFDUJOHUIJTTFSWJDFQSPHSBNZPVDBOWJTVBMJ[FUIFOVNCFS
of exposures made.
The following message will appear on the display:
EXPOSURE COUNTER
P=XXXXX C=XXXXX
where the number after the P indicates the panoramic exposure
made and the number after the C the teleradiographic ones.
F.4.3.13
Software version
With this service program you can check the software version
NPVOUFEPODPOWFSUFSCPBSE12POIPTUDPNQVUFSCPBSE1$
and on keyboard control board PC7.
The following message will appear on the display:
SOFTWARE VERSION
H=XX CV=XX CS=XX
where the number after H indicates the software version of the
host computer board PC1, after CV the software version of the
DPOWFSUFSCPBSE12BOEBGUFS$4UIFTPGUXBSFWFSTJPOPGUIF
keyboard control board PC7.
F.4.3.14
PA7/PH3 software version
This service program is applicable only for Orthoralix 9200 DDE
versions.
With this service program you can check the software version
mounted on the Auxiliary board PA7, on Panoramic DMU Interface
board PH3 and on Ceph DMU Interface board PH3.
The following message will appear on the display:
PA7/PH3s SW.VER.
A=XX HP=XX HC=XX
where:
a)
the number after A indicates the software version of the
Auxiliary board PA7
b)
the number after HP indicates the software version of the
Panoramic DMU Interface board PH3
c)
the number after HC indicates the software version of the
Ceph DMU Interface board PH3.
F.4.3.15
Ceph arm
This service program is applicable only for Orthoralix 9200 DDE
versions.
This service program indicates if the ceph arm has been properly
SFDPHOJ[FECZUIFIPTUDPNQVUFSCPBSE1$
The following message will appear on the display:
CEPH ARM Y/N
< NO >
XIFSF/0JTUIFEFGBVMUDPOEJUJPO
When the ceph arm is installed, at the first system switching
on, the host computer board PC1 is detecting the presence of
the ceph arm and automatically changes the setting from NO
to YES.
Please take note that, if the ceph arm is removed, in this service
program will remain the setting YES: this can generate error mes-
sages when the system is switched on. So, in this case, modify
manually the setting from YES to NO by pressing the keys projec-
tion+ or projection- on the keyboard.
F.4.3.16
X-ray enable
With this service program you can have a complete X-ray cycle
but without X-ray emission.
This could be useful for example during an exhibition, or in any
other case where you require a complete simulation but without
X-ray.
The following message will appear on the display:
XRAYS ENABLE Y/N
< YES >
XIFSF:&4JTUIFEFGBVMUDPOEJUJPO
1.
modify the condition (YES / NO) by pressing the keys projec-
tion+ or projection- on the keyboard.
2.
the new condition will be stored when the system is switched
off.
(03.0)E