ORTHORALIX 9200
Service Manual
F-21
Gendex Dental Systems 4519-190-71193-7
(04.0)E
F.3.56
HW fault 153
This alarm mesage is indicating an hardware fault in the PLD
registers of the auxiliary board PA7.
Press the RESET pushbutton on the control panel to cancel the
alarm.
If after the RESET the alarm is still present, replace the auxiliary
board PA7.
*GUIFBVYJMJBSZCPBSE1"OFFETUPCFSFQMBDFEUIF
*1BEESFTTBOEUIF4VCOFU.BTLNVTUCFTUPSFEJO
the new PA7 by using the Service Programs (see
the relevant paragraphs of this section for more
information).
F.3.57
HW fault 154
This alarm mesage is indicating an hardware fault in the internal
communication bus of the auxiliary board PA7.
Press the RESET pushbutton on the control panel to cancel the
alarm.
If after the RESET the alarm is still present, replace the auxiliary
board PA7.
*GUIFBVYJMJBSZCPBSE1"OFFETUPCFSFQMBDFEUIF
*1BEESFTTBOEUIF4VCOFU.BTLNVTUCFTUPSFEJO
the new PA7 by using the Service Programs (see
the relevant paragraphs of this section for more
information).
F.3.58
HW fault 181
This alarm mesage is indicating that a panoramic DMU is inserted
in the ceph bay.
Install the correct DMU and press the RESET pushbutton on the
control panel to cancel the alarm.
F.3.59
HW fault 200
This alarm mesage is indicating an hardware fault in the internal
timers of the auxiliary board PA7.
Press the RESET pushbutton on the control panel to cancel the
alarm.
If after the RESET the alarm is still present, replace the auxiliary
board PA7.
*GUIFBVYJMJBSZCPBSE1"OFFETUPCFSFQMBDFEUIF
*1BEESFTTBOEUIF4VCOFU.BTLNVTUCFTUPSFEJO
the new PA7 by using the Service Programs (see
the relevant paragraphs of this section for more
information).
F.3.60
Host unavailable
This alarm message is generated when in the menu Config (en-
abled by clicking on the Orthoralix 9200 DDE icon in the system
tray of Windows) is not done the correct configuration:
1.
in the multiroom mode is not selected a default Personal
Computer
2.
in the single Personal Computer connection, it was not
selected the unique Personal Computer.
To restore the proper working condition, do the correct selection
in the menu Config.
Press the RESET pushbutton on the control panel to cancel the
alarm.
F.3.61
Host busy
This alarm message is generated when the Personal Computer
connected to the Orthoralix 9200 DDE is not ready to acquire the
images from the Orthoralix.
This may happen when in the Personal Computer others software
applications are running or when the Orthoralix 9200 DDE activex
driver is not correctly installed.
Press the RESET pushbutton on the control panel to cancel the
alarm.