Chapter 4. Error Codes and Troubleshooting
76
PanaFlow HT User’s Manual
4.2.2 Communication Error String
The PanaFlow HT flowmeter electronics includes two independent sub-systems. The purpose of the Communication
error string is to convey to the operator an issue with communication between these two sub-systems.
Figure 28: Communications Error String
4.2.3 Flow Error String
Flow errors are errors detected by the UMPU (Ultrasonic Measurement Processing Unit) in the course of making a
flow measurement. These errors can be caused by disturbances in the fluid, such as excessive particles in the flow
stream or extreme temperature gradients. The errors could also be caused by an empty pipe or other such issue with the
fluid itself. Flow errors are typically not caused by a malfunction of the flow measurement device, but by an issue with
the fluid itself.
Table 4: Communication Error String
Error Header
Error Message
C1
UMPU Comm error
Table 5: Flow Error String
Error Header
Error Message
Explanation
E29
VelocityWarning
E29 indicates that the flow rate has exceeded the range
of the LWL (lower warning limit) or UWL (upper
warning limit). The purpose of this warning is to alert
the operator that the flow rate is approaching the LFL
(lower functional limit) or the UFL (upper functional
limit). At the Warning Limit threshold, the flowmeter
will continue to measure flow and drive the SIL
output, but if the flow rate reaches the Functional
Limit threshold the SIL output will go to the
Dangerous Detected state until an Authorized User
intervenes. E29 gives the operator a chance to correct
the situation before going to the DD state.
E22
SingleChAccuracy
E22 indicates that an error was detected on one of the
flow channels. This applies to multiple channel
systems only. For example, there may be an issue in
the channel 1 measurement but not the channel 2
measurement.
E23
MultiChAccuracy
E23 indicates an error on multiple channels.
Содержание PanaFlow HT
Страница 1: ...GE Measurement Control Flow 910 294U Rev A September 2012 PanaFlow HT User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Chapter 1 Introduction 4 PanaFlow HT User s Manual no content intended for this page ...
Страница 84: ...Chapter 3 Initial Setup and Programming 74 PanaFlow HT User s Manual no content intended for this page ...
Страница 98: ...88 PanaFlow HT User s Manual no content intended for this page ...
Страница 99: ...PanaFlow HT User s Manual 89 Appendix B Menu Maps Appendix B Menu Maps ...
Страница 100: ...Appendix B Menu Maps 90 PanaFlow HT User s Manual ...
Страница 137: ...128 PanaFlow HT User s Manual no content intended for this page ...
Страница 142: ...PanaFlow HT User s Manual 133 D 5 Review Menu Figure 39 Review Menu ...
Страница 143: ...134 PanaFlow HT User s Manual no content intended for this page ...
Страница 147: ...138 PanaFlow HT User s Manual no content intended for this page ...
Страница 153: ...144 PanaFlow HT User s Manual no content intended for this page ...
Страница 155: ...146 PanaFlow HT User s Manual no content intended for this page ...
Страница 161: ...Warranty 152 PanaFlow HT Ultrasonic Liquid Flowmeter User s Manual no content intended for this page ...
Страница 162: ......