![GE PanaFlow HT Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/ge/panaflow-ht/panaflow-ht_user-manual_128416031.webp)
PanaFlow HT User’s Manual
21
Chapter 2. Installation
2.5.6 Wiring the Line Power
The PanaFlow HT may be ordered for operation with power inputs of 100-240 VAC or 15-30 VDC. The label on the
side of the electronics enclosure lists the meter’s required line voltage and power rating. The fuse size is listed in
Appendix A,
Specifications
. Be sure to connect the meter to the specified line voltage only.
Note:
For compliance with the European Union’s Low Voltage Directive (2006/95/EC), this unit requires an external
power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such,
clearly visible, directly accessible, and located within 1.8 m (6 ft) of the PanaFlow HT.
Refer to Figure 10 on page 11 to locate the terminal blocks and connect the line power as follows:
WARNING! Improper connection of the line power leads or connecting the meter to the incorrect line
voltage may damage the unit. It may also result in hazardous voltages at the flowcell and
associated piping as well as within the electronics enclosure.
1.
Prepare the line power leads by trimming the line and neutral AC power leads (or the positive and negative DC
power leads) to a length 0.5 in. (1 cm) shorter than the ground lead. This ensures that the ground lead is the last to
detach if the power cable is forcibly disconnected from the meter.
2.
Install a suitable cable clamp in the conduit hole. If possible, avoid using the other conduit holes for this purpose, to
minimize any interference in the circuitry from the AC power line.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in
Appendix G,
CE Mark Compliance
.
3.
Route the cable through the conduit hole and connect the line power leads to the power terminal, using the pin
number assignments shown in Figure 10 on page 11.
4.
Leaving a bit of slack, secure the power line with the cable clamp.
5.
If the unit has been completed, reinstall the wiring cover on the enclosure and tighten the set screw.
WARNING! Make sure all covers, with their O-ring seals, are installed and the set screws tightened before
applying power in a hazardous environment.
CAUTION!
The transducers must be properly wired before applying power to the meter.
Содержание PanaFlow HT
Страница 1: ...GE Measurement Control Flow 910 294U Rev A September 2012 PanaFlow HT User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Chapter 1 Introduction 4 PanaFlow HT User s Manual no content intended for this page ...
Страница 84: ...Chapter 3 Initial Setup and Programming 74 PanaFlow HT User s Manual no content intended for this page ...
Страница 98: ...88 PanaFlow HT User s Manual no content intended for this page ...
Страница 99: ...PanaFlow HT User s Manual 89 Appendix B Menu Maps Appendix B Menu Maps ...
Страница 100: ...Appendix B Menu Maps 90 PanaFlow HT User s Manual ...
Страница 137: ...128 PanaFlow HT User s Manual no content intended for this page ...
Страница 142: ...PanaFlow HT User s Manual 133 D 5 Review Menu Figure 39 Review Menu ...
Страница 143: ...134 PanaFlow HT User s Manual no content intended for this page ...
Страница 147: ...138 PanaFlow HT User s Manual no content intended for this page ...
Страница 153: ...144 PanaFlow HT User s Manual no content intended for this page ...
Страница 155: ...146 PanaFlow HT User s Manual no content intended for this page ...
Страница 161: ...Warranty 152 PanaFlow HT Ultrasonic Liquid Flowmeter User s Manual no content intended for this page ...
Страница 162: ......