Appendix C. Modbus Map
104
PanaFlow HT User’s Manual
Register
(in Hex)
Register (in
Decimal)
Access
Level
Variable
Unit Group
Read Only/
Read Write
Format
219A
8602
Gen User
Output_1 Control Output Measurement
Register Address
18
RW
INT32
219C
8604
Gen User
Output_1 Control Output Unit code
18
RW
INT32
219E
8606
Gen User
Output_1 Test Control Output
18
RW
INT32
21A0
8608
Gen User
Output_1 Reserved
18
RW
INT32
21A2
8610
Gen User
Output_1 Test Mode
0: Test Off; 1: Test On
18
RW
INT32
21A4
8612
Gen User
Output_1 Pulse Measurement Register
Address
18
RW
INT32
Digital Out 2 Real Read Write
2400
9216
Gen User
Output_2 Pulse Value
18
RW
(IEEE 32 bit)
2402
9218
Gen User
Output_2 Pulse Time
10
RW
(IEEE 32 bit)
2404
9220
Gen User
Output_2 Frequency Base Value
18
RW
(IEEE 32 bit)
2406
9222
Gen User
Output_2 Frequency Full Value
18
RW
(IEEE 32 bit)
2408
9224
Gen User
Output_2 Alarm Value
18
RW
(IEEE 32 bit)
240A
9226
Gen User
Output_2 Control Output Value
18
RW
(IEEE 32 bit)
Digital Out 2 Integer Read Write
2500
9472
Gen User
Output_2 Test Pulse Value
18
RW
INT32
2502
9474
Gen User
Output_2 Frequency Full Frequency
6
RW
INT32
2504
9476
Gen User
Output_2 Test Frequency Value
6
RW
INT32
2506
9478
Gen User
Output_2 Frequency Error Handling
Value
6
RW
INT32
2580
9600
Gen User
Output_2 type
18
RW
INT32
2582
9602
Gen User
Output_2 Pulse Value Unit Code
18
RW
INT32
2584
9604
Gen User
Output_2 Pulse Error Handling
18
RW
INT32
2586
9606
Gen User
Output_2 Frequency Measurement
Register Address
18
RW
INT32
2588
9608
Gen User
Output_2 Frequency Error Handling
18
RW
INT32
258A
9610
Gen User
Output_2 Frequency Unit code
18
RW
INT32
258C
9612
Gen User
Output_2 Alarm State
18
RW
INT32
258E
9614
Gen User
Output_2 Alarm Type
18
RW
INT32
2590
9616
Gen User
Output_2 Alarm Measurement Register
Address
18
RW
INT32
2592
9618
Gen User
Output_2 Alarm Unit code
18
RW
INT32
2594
9620
Gen User
Output_2 Test Alarm
18
RW
INT32
2596
9622
Gen User
Output_2 Control Output State
18
RW
INT32
2598
9624
Gen User
Output_2 Control Output Type
18
RW
INT32
259A
9626
Gen User
Output_2 Control Output Measurement
Register Address
18
RW
INT32
259C
9628
Gen User
Output_2 Control Output Unit code
18
RW
INT32
259E
9630
Gen User
Output_2 Test Control Output
18
RW
INT32
25A0
9632
Gen User
Output_2 Phase Shift
18
RW
INT32
Table 10: XMIT Modbus Map (cont.)
Содержание PanaFlow HT
Страница 1: ...GE Measurement Control Flow 910 294U Rev A September 2012 PanaFlow HT User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Chapter 1 Introduction 4 PanaFlow HT User s Manual no content intended for this page ...
Страница 84: ...Chapter 3 Initial Setup and Programming 74 PanaFlow HT User s Manual no content intended for this page ...
Страница 98: ...88 PanaFlow HT User s Manual no content intended for this page ...
Страница 99: ...PanaFlow HT User s Manual 89 Appendix B Menu Maps Appendix B Menu Maps ...
Страница 100: ...Appendix B Menu Maps 90 PanaFlow HT User s Manual ...
Страница 137: ...128 PanaFlow HT User s Manual no content intended for this page ...
Страница 142: ...PanaFlow HT User s Manual 133 D 5 Review Menu Figure 39 Review Menu ...
Страница 143: ...134 PanaFlow HT User s Manual no content intended for this page ...
Страница 147: ...138 PanaFlow HT User s Manual no content intended for this page ...
Страница 153: ...144 PanaFlow HT User s Manual no content intended for this page ...
Страница 155: ...146 PanaFlow HT User s Manual no content intended for this page ...
Страница 161: ...Warranty 152 PanaFlow HT Ultrasonic Liquid Flowmeter User s Manual no content intended for this page ...
Страница 162: ......