Chapter 2 Equipment Requirements
1 System Components
1.1 Presentation of the 3 Rooms
1.1.1 Examination room
•
Gantry
•
Omega Patient Table
•
Performix X–Ray Tube Assembly
•
Collimator
•
Digital Detector
•
TSUI
•
ITU
WARNING
THE ELECTRONIC CABINETS (SYSTEM CONTROL CABINET, OPTIONAL LD
CABINET, COOLIX 4100 CHILLER, DETECTOR CONDITIONER AND FLUORO
UPS WHEN INSTALLED) INCLUDE FANS THAT ARE CREATING AIR-
CIRCULATION OF PULSED-AIR. WHEN THIS PULSED AIR IS IN AN
ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF
THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS
OR CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
1.1.2 Technical room
•
System Control Cabinet
•
Coolix 4100 chiller
•
Detector Conditioner
•
Fluoro UPS (option) (1 Cabinet)
•
Fluoro UPS IF box (option)
•
PDB
•
EMC Filter (Option)
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
43