1.2.3 User Interfaces
Illustration 2-3: User interfaces with Omega table
1.2.4 Description of the X–Ray generator
The System uses a High Voltage System with component parts as follows according to the
power requested (100 kW):
•
Jedi – System Control Cabinet,
•
Ingrid H.V. tank mounted on the Performix 160A X–Ray tube housing.
1.2.5 Description of the X–Ray head
The Optima System uses a Performix 160A X–Ray tube housing:
•
X–Ray tube
•
collimator
•
Ingrid HV tank
•
oil/water exchanger
•
contour filter
•
tube chiller
NOTE:
An external recirculating chiller is mandatory.
It is mandatory to place the Chiller in the Technical Room.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
47