1.2 Power and grounding
The power cords between the peripherals in the control room and the power strip are not shown
on the following diagram. All are single phase power cords (with phase, neutral line and
protective Earth) and are provided by GE.
1.2.1 Schematic
Illustration 5-1: Electrical and grounding schematic
NOTE:
For the complete system connection schematic, refer to the MIS Map and MIS Chart
provided in the Advanced Service Manual.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 5 Electrical Requirements
165