Seismic attachment hardware shown on seismic calculations may differ from hardware supplied
with system. Any additional hardware that is required will be the responsibility of the institution
and/or their contractor. Contact your local GE Installation Program Manager to obtain seismic
calculations.
Seismic calculations are per California Building Code (CBC) and International Building Code
(IBC).
The following shows center-of-gravity information for system components:
•
System Control Cabinet,
•
•
•
•
PDB CE,
•
•
Detector Conditioner,
•
Omega IV Table,
•
•
•
19” Desk Mounted Monitor,
•
Mavig Overhead Flat Panel Suspension,
•
Ergo Overhead Flat Panel Suspension,
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
137
Содержание Optima IGS 320