However, when installed and operated as specified herein, the system will maintain its essential
performance by continuing to acquire, display, and store diagnostic quality images safely.
For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3
) apart from the system (in order to avoid images interferences risks).
2.5 Use Limitation
WARNING
THE USE OF ACCESSORIES, TRANSDUCERS, AND CABLES OTHER THAN
THOSE SPECIFIED MAY RESULT IN DEGRADED ELECTROMAGNETIC
COMPATIBILITY OF THE SYSTEM.
In case of an Optima system, the Gantry (digital detector) shall be apart 1 meter from the
generator cabinet, and 1 meter apart from the analog (CRT) monitors. These distances
specifications shall minimize the low frequency magnetic field interference risk.
Other electrical equipment may disturb and interfere with these Optima components. The
control of the clearing distances from the noise sources is recommended from the HF
electrosurgery generator, power supplies converters from nearby monitors or from other close
electrical equipment). Refer to respective device manufacturers instructions & recommendations
in such cases.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
152