The relative humidity shall be at least 30 percent.
○
The dissipative material shall be connected to the system ground
reference, if applicable.
NOTICE
Route separately EMI filter incoming power lines and outgoing power lines (refer
to
Electrical Connections).
NOTE:
The maximum distance between the EMI filter and the PDB is 3 m.
NOTE:
The Fluoro UPS (CE) option requires an EMI filter box placed upstream the PDB. The
EMI filter box will be delivered as a single product by GE Consumer & Industrial.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
154