2.3 Cable Channeling
2.3.1 General
High voltage and power cables must be separated from other cables. Use a separate trough in
the duct system, or use a separate conduit. Minimize cable length between the line disconnect
and the System Cabinet power unit to reduce voltage regulation problems and wiring costs.
For information about the cables supplied with your system, please refer to
2.3.2 Conduit
Separate conduits must be used for power and signal wires. These wires must be kept
separated from each other.
Using conduit imposes some important considerations when used with this system. Of primary
concern, the majority of cables used are pre–terminated. Pre–termination greatly simplifies
interconnection but makes cable–pulling difficult because of the added dimensions of the
connectors.
Conduit must be large enough to pass the cable and connector through with all other cables
already in the conduit. Also, the size of conduit chosen must allow for future growth. There is the
possibility of additional cables being added later as the system is developed and options are
added.
The use of conduit is recommended for cables running overhead between rooms, especially
when a diagonal run provides the shortest cable path
Separate conduits must be used for Hospital and Fluoro UPS power wires. These wires must be
kept separated from each other (at least 30 cm).
2.3.3 Electrical Ducts
It’s important that electrical ducts have separate compartments for power and signal wires.
These wires must be kept separated from each other for proper system operation.
Electrical ducts have advantages, when used with a single room or two adjacent rooms.
Electrical ducts combine cabling in a neat and functional appearance, with accessibility and
room for expansion.
NOTE:
Medrad AVANTA and Mac-lab cables exit behind the table in the patient room.
NOTE:
It is the responsibility of the site planner to provide the appropriate solution to the table exit (e.g
gas box, Clab II, Tram module, connection interface box)
NOTE:
MEDRAD Avanta Table mount: A 76.2 mm (3 in) and max 25 m (984 in) length
conduit between technical room and patient room shall be prepared below the floor for
the three injector cables. It is recommended to use the MEDRAD Avanta floor
mounting bracket to cover the duct hole in the patient room if there is no gases box.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
186
2 System Cable Information
Содержание Optima IGS 320
Страница 1: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual OPERATING DOCUMENTATION 5537562 1 1EN Revision 3...
Страница 2: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 2...
Страница 70: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 70 1 System Components...
Страница 201: ......