NOTE:
(2) Special Humidity Instructions: The following parts can be shipped in standard
shipment conditions with the requirement that on arrival to installation site, and before
supplying power to these parts, they shall be kept in an environmental relative
humidity equal or lower than their specified capability, and that’s for a minimum of 48
hours.
•
CVI Injector (Maximum specified relative humidity capability = 85%)
•
Diamentor (Maximum specified relative humidity capability = 80%)
•
1MP MX191 LCD Monitor (Maximum specified relative humidity capability = 80%)
4.2 Fluoro UPS Shipment Duration
NOTE:
In the case the material shipped will be subject to different temperature values during
shipment, the maximum shipment duration can be calculated, depending on the
duration of different temperature values application, and using values given in the
tables above
4.2.1 Fluoro UPS (CE and UL)
The maximum shipment duration is given in the table for both Fluoro UPS CE and UL. It
depends on the temperature to which the material will be subject during shipment. This
limitation is due to the capability of batteries included in UPS.
Table 1-5: Maximum shipment duration for Fluoro UPS
Temperature
Shipment max duration (Weeks)
55°C (131°F)
2
50°C (122F°)
3
40°C (104°F)
6
30°C (86°F)
12
4.3 Handling instructions
The packaging of the following components must be marked with special handling instructions
for transport and storage.
•
System Control Cabinet:
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
40
4 Product Storage and Handling Requirements
Содержание Optima IGS 320
Страница 1: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual OPERATING DOCUMENTATION 5537562 1 1EN Revision 3...
Страница 2: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 2...
Страница 70: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 70 1 System Components...
Страница 201: ......