1.1.2 Options
Table 5-2:
Option
Nominal voltage
Frequency
Nominal Power consumption Type of power input
AW
100-127 V / 200-240 V
50 Hz or 60 Hz
11 A / 5.5 A
Single phase
WARNING
PRIOR TO EACH INSTALLATION, ENSURE THAT THE ECG POWER CABLE
IS CONNECTED TO A LINE THAT IS PROTECTED AGAINST SHORT
CIRCUIT HAZARDS, ACCORDING TO LOCAL REGULATIONS AND THAT
THE ECG MONITOR IS POWERED ACCORDING TO IEC60601-1
REQUIREMENTS (LEAKAGE CURRENT AND GROUNDING).
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
164
1 Power Requirements