Height:
min.: 300 mm (12”)
max.: 350 mm (14")
Depth:
min.: 30 mm (2")
max.: 120 mm (4")
Weight:
min.: 5 kg (11 lbs)
max.: 7 kg (15.4 lbs)
Power:
No external power supply
Electrical:
IEC 320 connector connected to Optima system.
Ground:
Screw-on connector
Video:
VGA connector
Fixing:
VESA 100 mm x 100 mm (4" x 4") connection, in the middle on the backside
1.2.8.3 System Cable Information
GE Healthcare provides 24 m (78'9") system cables for suspension. The length of cables
dedicated on the rail cable track is recommended as 3.8 m (12'6"). After leaving cable track,
system cables go to technical room through the conduit on the ceiling, ECG monitor VGA and
power cable go to the EKG system through another conduit. From entrance of System Control
Cabinet to their correspondent connectors, the length of the cables shall be 1.3 m (4’3’’).
Illustration 2-10: Ergo suspension system cable routing scheme
1.2.9 Description of the DL and associated devices
An Optima System uses the DL Digital Imaging system.
•
The System Control Cabinet contains:
1. DL Computer,
2. Firewall
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
53