2.2 Room Layout Drawings
2.2.1 Patient Room Layout
2.2.1.1 Room Layout for Optima System (examples of room layouts)
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS
MANDATORY.
THE ELECTRONIC CABINETS (SYSTEM CONTROL CABINET, COOLIX 4100
CHILLER, DETECTOR CHILLER AND FLUORO UPS WHEN INSTALLED )
INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF PULSED–AIR.
WHEN THIS PULSED AIR IS IN AN ENVIRONMENT THAT MAY CONTAIN
AIRBORNE PATHOGENS LIKE AN EXAM ROOM/CONTROL ROOM, THERE
IS A RISK OF TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM
PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL (NOSOCOMIAL
DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
and
.
1. Gantry
2. Patient table
3. XT monitor suspension
4. Monitor frame
5. Injector on Pedestal
NOTE:
The LAN outlet (RJ45 connector) must be located less than 1 meter (3 feet) from the
System Control Cabinet (3).
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
89