Following is a list of options in the boot code service menu.
Boot Loader Service Menu Options
Menu Option
Description
CHANGE INTERNET
ADDRESS
This option is used when changing the internet
protocol (IP) address.
WARNING
Duplication of an Internet address on a network
causes data loss and possible Unity Network
problems. If you change the factory assigned Internet
address, you must record all other Internet addresses
used on your network to avoid duplication.
SHOW INSTALL OPTIONS
This option lists the options installed on the monitor.
CLEAR CONFIGURATION
MEMORY
This option erases all patient histories and
programmed defaults in the monitor’s protected
memory. Clearing the configuration memory returns
the monitor defaults to the factory installed defaults.
SET CONFIGURATION
This menu contains options for configuring the
monitor.
Refer to
VIDEO TEST SCREENS
Various color screens for testing the display.
BATTERY SIMULATION
This option is for engineering use only.
WAKE UP BATTERY
This option is used when the battery is dead. Refer to
“Battery Messages Displayed in the Battery Fuel
Gauge Icon” on page 5-13
.
OPTIONS MENU
A unique password is required for each option.
Contact your sales/service representative to obtain a
password. You must provide your product serial
number and Ethernet address. (The Ethernet address
is displayed in the Boot Code banner information.)
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......