Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Appendix A – Electromagnetic Compatibility (EMC):
Electromagnetic Compatibility (EMC)
Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
The Dash 3000/4000 Patient Monitor is intended for use in the
electromagnetic environment specified below. It is the responsibility of
the customer or user to assure that the Dash 3000/4000 Patient Monitor
is used in such an environment.
NOTE
:
U
t
is the AC mains voltage prior to application of the test level.
Immunity Test
EN 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Electrostatic
Discharge (ESD)
EN 61000-4-2
± 6 kV contact
± 6 kV contact
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
± 8 kV air
± 8 kV air
Electrical Fast
Transient/Burst
EN 61000-4-4
± 2 kV for power supply lines
± 2 kV for power supply
lines
Mains power should be that of a typical commercial
or hospital environment.
±1 kV for input/output lines
±1 kV for input/output
lines
Surge
EN 61000-4-5
± 1 kV differential mode
± 1 kV differential mode
Mains power should be that of a typical commercial
or hospital environment.
± 2 kV common mode
± 2 kV common mode
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN 61000-4-11
<40% U
t
(>60% dip in U
t
)
for 5 cycles
<40% U
t
(>60% dip in
U
t
) for 5 cycles
Mains power should be that of a typical commercial
or hospital environment. If the user of the equipment
requires continued operation during power mains
interruptions, it is recommended that the equipment
be powered from an uninterruptible power supply or
a battery.
<70% U
t
(>30% dip in U
t
)
for 25 cycles
<70% U
t
(>30% dip in
U
t
) for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 s
<5% U
t
(>95% dip in U
t
)
for 5 s
Power Frequency
(50/60 Hz)
Magnetic Field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......