Revision A
Dash 3000/4000 Patient Monitor
4-35
2000966-339
Maintenance:
Checkout Procedures
Identify the Patient Parameter and Software Options Enabled on the Monitor
The check out procedures support all enabled patient parameter and
software options. Complete the following procedure to record the options
enabled on the monitor.
1. From the Main Menu, select
MORE MENUS
>
MONITOR SETUP
>
SOFTWARE CONFIGURATION
.
2. Record the software options that are enabled on the monitor.
3. Complete the checkout procedures for the monitor.
Monitor Power-up Tests
NOTE
When the monitor is connected to a docking station, use the docking
station’s power cable instead of the monitor’s power cable.
1. Remove the batteries and unplug the monitor (or the docking station)
from AC power to turn the monitor off.
2. Restore the batteries to the monitor and plug the monitor (or the
docking station) into AC power to turn the monitor on.
3. Verify all of the front panel indicators illuminate on power up.
4. Verify the AC indicator on the monitor stays illuminated.
NOTE
If the AC LED stays on, but the screen is blank, the monitor is
likely in “standby mode” (battery charging). Press the
POWER
button to enter the normal mode.
If the AC indicator is on, continue with the tests.
Option
X
Comments
0BP (invasive)
No invasive BP options are enabled. As a result, no
invasive blood pressure test is required.
1BP or 2BP
(invasive)
Complete the
“Invasive Blood Pressure Tests”
3BP or 4BP
(invasive)
Complete the
“Invasive Blood Pressure Tests”
44. This test requires the Dual BP cable to complete the
checkout procedure. See
12SL Analysis
Complete the
39. If 12SL is not enabled, complete the
ETCO
2
(CO
2
)
(End-tidal CO
2
)
Complete
Network
Connectivity
Complete the
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......