Configuration:
Advanced User Procedures
Copy Stored Monitor Defaults
WARNING
Loss of Defaults. Copying monitor defaults from the
server monitor erases all of the monitor defaults on the
client monitor.
To copy monitor defaults from the server monitor, the client monitor
must be in the Discharged state. If the client monitor is set to the OR
patient-monitor type, it must not be in Combo or Rover Combo Admit
mode.
NOTE
Make sure that the client monitor shares the same configuration as
the server monitor. Refer to
“Transfer Monitor Defaults” on page 6-
.
1. At the client monitor, start from the Main Menu and select
MORE
MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the
Trim Knob
control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407)
3. Select
COPY UNIT DEFAULTS
.
4. Select
SELECT UNIT TO COPY MONITOR DEFAULTS FROM
.
5. From the list of units on the Unity network, choose the unit the
server monitor is in.
6. Select
SELECT BED TO COPY MONITOR DEFAULTS FROM
.
7. Scroll through the list of beds within the selected unit until the
desired server monitor is found. Select it, and answer
YES
to the
confirmation popup menu.
NOTE
After copying monitor defaults, the first set of defaults is
automatically activated. If another set of defaults is desired, the user
must manually select it from the Recall Defaults menu.
After copying monitor defaults from another bed (the server monitor),
verify that the defaults were transferred and arrhythmia levels are as
desired.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......