Revision A
Dash 3000/4000 Patient Monitor
9-11
2000966-339
Assembly Drawings:
Theory Of Operation
The DAS system block diagram with Masimo SET SPO2 and four
SuperStat BP channels consists of the following three sections.
Parameter Functions of the DAS with Masimo SET SPO2
and Four SuperStat BP Channels
Except for the SPO2 parameter, the DAS with the Masimo SET SPO2
parameter supports the same parameters as DAS with the GE SPO2
parameter. Refer to
.
Probes and Cable
The Masimo compatible DAS is intended for use exclusively with pulse-
oximetry probes from Masimo Corporation. An adapter cable is used to
interface the probes to the DAS front panel Nicolay connector. The
Nicolay connector utilizes a key pattern unique to Masimo.
Processor/Power Management Subsystem
Overview
The main processor/power management PCB contains the electrical
hardware to provide data processing and display of patient and monitor
configuration data, communication and interface circuitry, and power
conversion and battery management functions for the monitor.
The high level of integration attained in the design of the processor/
power management PCB is attributed to the use of several highly
integrated devices. A complex communications controller, ASIC, and
battery management hardware significantly improve the performance
and reduce the complexity and cost of the assembly. In addition to the
CPU, the main microcontroller contains a six-channel communications
processor as well as memory, PC Card, and video controllers. The devices
used in the core processing architecture all operate at 3.3 V to minimize
power consumption, yet the main processor and ASIC are tolerant of 5 V
hardware peripherals.
Section
Description
ECG/Respiration
Separately isolated section includes a 7.3728MHz Motorola
MC68HSC705C8A microcontroller, A/D conversion, signal
processing hybrids and DC-DC isolation converter.
Main DAS
Separately isolated section includes a 22.1184MHz Motorola
MC68332 microcontroller with FLASH and SRAM memory, an
Analog Devices ADSP-21062 32-bit floating point digital signal
processor running at 12.096MHz, A/D conversion, signal
processing hybrids/modules and DC-DC isolation converter.
Non-Isolated
circuits
Includes ECG and Main DAS isolation power conversion circuitry,
serial and parallel host interfaces and NBP pump, valves, and
over-pressure circuitry.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......