Revision A
Dash 3000/4000 Patient Monitor
4-7
2000966-339
Maintenance:
Battery Power
Battery Power
Introduction
The monitor is designed to operate on battery power during transport or
whenever AC power is interrupted. A complete battery management
system allows you to obtain maximum battery performance. Audible and
visual alarms alert you when loss of power is imminent and on-screen
capacity gauges indicate battery charge condition and capacity.
Depending on usage, you can get up to 2.5 hours of run time per battery
for new, fully-charged batteries. Monitoring NBP, CO2 and SpO2 will
drain battery power faster than other parameters. The display
technology and brightness, graphing, and connection to a wireless
network can also alter battery run times.
NOTE
A
“BATTERY LOW”
message at the top of the screen warns you prior
to complete loss of battery power. You should replace the battery or
connect the monitor to an AC power source when the message is
displayed.
Exchangeable Battery Packs
The monitor uses two exchangeable battery packs. These battery packs
are a purchasable option. One battery pack can easily be exchanged
while the monitor operates from the other.
CAUTION
Leaks from battery cells can occur under extreme
conditions. The liquid is caustic to eyes and skin. If the
liquid comes into contact with eyes or skin, flush with
clean water and seek medical attention.
NOTE:
For optimal performance and safety, use only batteries supplied
by GE.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......