6-18
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Configuration:
Advanced User Procedures
Troubleshooting Monitor Defaults Transfer
Below is a list of error messages that may display when performing a
monitor defaults transfer.
Storing Monitor Defaults
Copying Stored Monitor Defaults
Message
Description
ERROR - UNIT DEFAULTS HAVE NOT
BEEN SAVED
Monitor defaults were not saved on the
server monitor due to a memory error in the
monitor.
Message
Description
ERROR COPYING UNIT DEFAULTS -
NETWORK ERROR
Either an error has occurred while the
defaults were being transferred, or defaults
have not been saved on the server monitor.
ERROR COPYING UNIT DEFAULTS -
UNIT TYPE MISMATCH
The server and client monitors are set to
different unit types (i.e. ADULT, NEO, or OR)
ERROR COPYING UNIT DEFAULTS -
SOFTWARE VERSION MISMATCH
The server and client monitors have different
software versions (i.e. V2C, V3A).
ERROR COPYING UNIT DEFAULTS -
DATA CORRUPTION ERROR
The monitor defaults transferred have
become corrupted, possibly due to a
memory error.
ERROR COPYING UNIT DEFAULTS -
SOFTWARE FEATURE LEVEL
MISMATCH
The server and client monitors are set to
different software levels (i.e. Basic, Cardiac,
Cardiopulmonary).
ERROR COPYING UNIT DEFAULTS -
COUNTRY CODE MISMATCH
The server and client monitors are set to
different country codes (i.e.
DEFAULT
or
FRANCE
).
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......