Revision A
Dash 3000/4000 Patient Monitor
8-17
2000966-339
Field Replaceable Units and Upgrades:
Display Assembly
Replacing the Backlight Inverter PCB
Complete the following steps for the monitor you are servicing.
Dash 3000 monitor
with display shield flex circuit
Dash 4000 monitor with Sharp display
assembly
1. Remove the handle assembly. Refer to page 8-7.
2. Remove the display assembly from the main unit. Refer to page 8-9.
3. Remove and replace the backlight inverter. Follow the steps for the monitor you are
servicing.
a. Flip over the display shield flex circuit.
Remove the male end of the two snap
rivets.
b. Flip over the display shield flex circuit.
Disconnect the display shield cable
from the backlight inverter PCB.
c.
Use a screw driver to remove the
female end of the two snap rivets. Do
not re-use these rivets.
a. Open the display assembly. Refer to
b. Disconnect the cable connectors from
the backlight inverter PCB
compression fittings.
c.
Remove the two screws anchoring the
backlight inverter PCB to the display
mounting plate.
4. Replace the defective backlight inverter PCB.
5. Reassemble the display assembly in reverse order.
plastic snap rivets (male end)
500A
display shield cable
plastic snap rivets (female end)
503A
882A
881A
881A
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......