Revision A
Dash 3000/4000 Patient Monitor
7-5
2000966-339
Calibration:
NBP, Analog Output ECG, Analog Output BP, and End-tidal CO2 Software Calibration
The table below lists items for connecting the NBP tube between the
manometer and NBP cuff:
WARNING
When the NBP cuff is used in this procedure, it must be
tightly wrapped around a rigid cylinder or pipe. DO NOT
put the NBP cuff around a human arm during the
calibration procedures due to the potential for injury.
Calibration Procedure
Using the
Trim Knob
control, access the Service Mode menu starting
from the Main Menu.
1. Select
MORE MENUS-> MONITOR SETUP-> SERVICE MODE.
2. Enter password using the
Trim Knob
control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407).
3. Select
CALIBRATE-> CALIBRATE NBP-> CAL ZERO OFF->
START
.
4. The text on the menu item changes from Cal Zero Off to Cal Zero In
Progress.
When the process is complete, the menu item shows that it is
OFF
again.
3
Gain Calibration Test
NOTE
To proceed with the Gain Calibration Test, setup the monitor and
test equipment following the guidelines illustrated in the NBP
Checkout Procedure, Chapter 4, Maintenance.
1. Connect a cuff and manometer to the monitor.
2. Turn the digital manometer on and adjust the range switch to 1000
mmHg.
3.
Select
CAL GAIN OFF-> CAL GAIN OFF->
START
.
Description
Part Number
Qty
NBP cuff coupling
400787-001
1
NBP hose coupling
46100-002
1
NBP tee
4745-101
1
NBP tubing
401582-001
2
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......