Revision A
Dash 3000/4000 Patient Monitor
6-7
2000966-339
Configuration:
Main Menu Selections
Set Graph Locations
Access Manual Graph Location option, starting from the Main Menu.
1. Select
MORE MENUS > MONITOR SETUP > GRAPH SETUP >
GRAPH LOCATION > MANUAL GRAPH LOCATION
.
2. Using the
Trim Knob
control, choose the manual graph location
from the list.
3. Select
ALARM GRAPH LOCATION
.
4. Using the
Trim Knob
control, choose the alarm graph location from
the list.
5. Select
PRINT WINDOW LOCATION
.
6. Using the
Trim Knob
control, choose the print window location from
the list.
7. Select
12 LEAD PRINT LOCATION
.
8. Using the
Trim Knob
control, choose the 12 lead print location from
the list.
Communication Confirmation
Confirm communication across the network.
1. Admit and generate a waveform at the monitor with a simulator.
2. Press
GRAPH GO/STOP
and observe graph output at chosen
locations.
Problems?
If the writer or printer does not graph:
Ensure the writer or printer is turned ON.
Check all cables for a good connection.
Check programmed alarms and manual graph locations at the
monitor.
If you do not have a waveform at the central station:
Ensure the central station software is compatible.
Check all cables for a good connection.
Check the programmed alarms and manual graph locations at the
monitor.
Ensure the care unit name is the same in the monitor and in the
central station.
Ensure the internet protocol (IP) addresses are programmed
correctly.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......