Revision A
Dash 3000/4000 Patient Monitor
4-43
2000966-339
Maintenance:
Checkout Procedures
5. Set the CO simulator to output injectate temperature (IT) readings,
as found in the following table:
6. Verify correct IT readings appear on the monitor display, as shown in
the table above.
7. Disconnect the CO cable adaptor from the
TEMP/CO
connector of the
monitor. This completes the CO tests.
Test with Marq II or Marq III Simulator
1. Connect the cardiac output cable adapter to the
TEMP/CO
connector
of the Tram module.
2. Connect the cardiac output adapter to the cardiac output cable
adapter and to the Marq II or to the MarqIII simulator.
3. At the monitor, select
MORE MENUS
>
CO
.
4. Set up the monitor:
a. Set
AUTO MODE:
to
ON
b. Set
INJECT TEMP:
to
BATH
c. Set
SIZE:
to
7
d. Set
INJECT VOL:
to
10CC
e. Set
COMPUTATIONAL CONSTANT:
to
0.540
5. Set the adapter to
0°
.
6. Turn the simulator ON. Select cardiac output menu corresponding to
the injectate temperature. (i.e. menu #92 for 0°, or #95 for 24°)
7. When the monitor screen displays the message,
INJECT WHEN
READY
, press
Enter
on the simulator.
8. When computing is complete, the CO reading should be
approximately 37°C at 5 liters of blood per minute
±
5%.
9. Repeat this test with the adapter set to
24
° and the
COMPUTATIONAL CONSTANT:
set to 0.595.
10. The output should be the same.
Simulator IT Setting
Monitor IT Reading Range
0.0°C
–0.3 – +0.3
8.0°C
7.7 – 8.3
15.0°C
14.7 – 15.3
24.0°C
23.7 – 24.3
29.6°C
29.3 – 29.9
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......