Maintenance:
Battery Maintenance
Conditioning the Battery With a Cadex SMart Two+ Charger
A conditioning cycle using the battery charger requires approximately
nine hours to complete. Complete the following steps to automatically
condition the battery.
1. Insert the battery into the battery charger.
2. Press the
CONDITION
button when one of the following conditions
occur:
While the
RUN
LED light is still flashing.
While the
CONDITION
LED is flashing.
3. Remove the battery from the battery charger when the
RUN
LED
illuminates. This completes the conditioning cycle.
Conditioning the Battery Inside a Monitor
WARNING
PATIENT RISK HAZARD — Never condition a battery
while the monitor is connected to a patient. Serious
injury or death could result.
Complete the following steps to manually condition the battery using the
monitor:
1. Remove all batteries from the monitor.
2. Insert one battery into Slot A of the monitor.
3. Disconnect the monitor from the AC power source.
4. Access the monitor’s Battery Status window to monitor the battery’s
Charge Level. See
“Identifying Battery Status Using the Battery
5. Allow the battery to discharge to less than a 90% Charge Level.
6. Re-connect the monitor to the AC power source and fully charge the
battery.
7. Disconnect the monitor from the AC power source and allow the
monitor to run until it displays the
BATTERY LOW
message, or
until the monitor shuts down.
8. Re-connect the monitor to the AC power source. Allow the battery to
fully charge to complete the conditioning cycle.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......